Home›Products›Liquid Flow Calibration Device›DN15–50 Fully Automatic Water Flow Calibration Device (Reference Meter Method)

DN15–50 Fully Automatic Water Flow Calibration Device (Reference Meter Method)

Request Specifications

Tell us the meter size, test method and required testing capacity.

DN15–50 water flow calibration device, reference meter, fully automatic. Request configuration details from KINMY.

Nominal Diameter
DN15–50
Test Method
Reference Meter Method
Automation
Fully Automatic

Overview

DN15–50 Fully Automatic Water Flow Calibration Device (Reference Meter Method) is listed in the KINMY Liquid Flow Calibration Device range with a configuration identified by DN15–50, Reference Meter Method, Fully Automatic. Contact KINMY for a project-specific specification sheet and quotation.

Configuration

Nominal Diameter DN15–50
Test Method Reference Meter Method
Automation Fully Automatic

Measurement and Test Methods

Measurement principles Reference Meter Method
Reference flow range 0.010–40 m³/h
Measurement modes Start-Stop Method, Flow-Time Method

Reference Standards and Instruments

Reference equipment listed for this catalog configuration. The supplied instrument set is confirmed with the project specification.

Reference meters Electromagnetic reference flowmeters

Reference Specifications

Measurement Uncertainty Urel ≤ 0.05% (k=2)
Flow Stability 1.0% for Start-Stop Method; 0.2% for Other Methods
Calibration Medium Pure Tap Water
Operating Humidity ≤ 93% RH
Operating Water Temperature 20℃ ± 10℃
Device Operating Temperature 4℃-50℃
Water Source Operating Pressure 0.0-0.5 MPa (Adjustable)
Air Source Operating Pressure 0.4-0.7 MPa
Number of Meters Under Test DN15: 10 units | DN20: 8 units | DN25: 6 units | DN32/40/50: 1 unit | Vertical Water Meters (DN15, DN20, DN25): 6 units | Vertical Water Meters (DN32): 2 units | Vertical Water Meters (DN40): 2 units | Vertical Water Meters (DN50): 1 unit
Meter Clamping Mechanism Dual-Cylinder External-Push Tandem Meter Clamping Device; Single Pneumatic Clamping Device for 10 Front 5 Rear Positions
Vacuum System Differential-Pressure Vacuum Generator, Vacuum Priming, Maintains Water Meters in Full-Water State
Supported Test Modes Fully Automatic, Semi-Automatic, Touchscreen Manual

Utilities and System Configuration

Water pump configuration Nanfang CHM10-50, Flow Rate: 8 m³/h, Head: 48 m, 1 unit | TD50-40G/2, Flow Rate: 50 m³/h, Head: 30 m, 1 unit
Water tank capacity 2000 L
Flow diverter DN25, DN50 Cross-Type Pneumatic Open Reversers, Low Noise & High Sensitivity, Stroke Difference 10 ms, Uncertainty ≤ 0.002%
Timing system (10-100) kHz (Adjustable); Accuracy Better Than 10⁻⁶
Pulse acquisition signal Pulse, f = (0~10) kHz
Data acquisition hardware 10 Mobile Phones (Camera Resolution ≥ 40 Million Pixels)
Tightness test pressure range (0~2.0) MPa